4-27
• Audio carrier frequency and level (numerical)
• Histogram graph of carrier levels
• Delta between audio and video levels
• Selected channel plan
• Test point compensation (appears only if a nonzero value
is programmed during setup)
• Limits annunciators
When the SCAN key is pressed a graph showing all carrier levels is
displayed. Use the RIGHT and LEFT arrow keys to position the
marker to the desired channel. The channel numbers can also be
entered directly using the numeric entry keys.
The UP and DOWN arrow keys can be used to adjust the reference
level of the graph.
Level Adjustments
Press the LVL soft key to edit the scale and reference level of the
Scan display. To adjust the Scale, use the UP and DOWN arrow keys
to select a scale that will provide the best viewing of the carrier
levels. To adjust the Ref Level, press the Ref Level soft key. Now
the Ref Level can be changed by using the UP and DOWN arrow
keys or by entering a numeric value followed by the ENTER key.
The reference level is the top line of the graph.
NOTE: The reference value is limited by unit and the scale setting.
Pressing the Auto Scale soft key will automatically set the reference
level for an optimum Scan display. The FCN + Scale key can also be
used to automatically set the reference level.
Frequency Adjustments
Press the FRQ soft key to edit frequency range parameters of the
Scan display. Use the UP and DOWN arrow keys or the numeric
entry keypad to set the start frequency. Press the Stop soft key to
adjust the stop frequency. The Full Span soft key can be used to set
the start and stop frequency to the full range of the unit.