User Manual 301/302
Installation
The instrument is fully solid state and dissipates only a small amount
of power. No special cooling is required. However, the instrument
should not be operated where the ambient temperature exceeds 40°C,
when the relative humidity exceeds 80% or condensation appears
anywhere on the instrument.
Avoid operating the instrument close to strong magnetic fields, which
may be found near high power equipment such as motors, pumps,
solenoids or high power cables. Use care when rack mounting to
locate the instrument away from sources of excessive heat or
magnetic fields. Always leave 4 cm (1.5 inches) of ventilation space
on all sides of the instrument.
Bench Operation
The Model 301/302 is shipped with plastic feet, tilt stand in place and
ready for use as a bench or portable instrument
Rack Mounting
The instrument may be rack mounted in a standard 19 inch rack. Rack
mounting ears are part of the 302 case. Telescopic rack mounting
slides are also available.
Configuring The Instrument
16