Figure 3.22, Sequence Table Download Format
There are a few things to consider when you prepare sequence tables
for the fast download format:
1. Data for each step is fixed and is comprised of 5 bytes
2.
The number of bytes in a complete sequence table must divide by
6. The Model 301/302 has no control over data sent to its
sequence table buffer during binary block downloads. Therefore,
wrong data and/or incorrect number of bytes will cause erroneous
sequence setting.
Using WaveWiz to
Generate
Sequences
Downloading
Segments
Segments are built and downloaded in turn to the 301/302 using
WaveWiz. Valid Waveforms can be waveforms using any of the
previously discussed methods. Prior to downloading the waveform, it
is necessary to set the segment number to where the waveform will be
sent. This is done using the segment control window which is turned
on by selecting “view segment window” in the view pull down menu.
Select the segment number by entering that number in the box, then
clicking on the Sel button.
Once the required waveform is defined, download it to the selected
segment by clicking on the
button on the toolbar. Repeat this
process for every segment required.
Selecting Number
of Sequences
In WaveWiz, use the sequence button on the sequence tab in the
Instrument Control Panel. The sequence table pop-up menu appears.
Configuring The Instrument
62